CT advantages of potential use of polymer plastic clips in neurocranium

Authors
  • Samir Delibegović orcid logo (University of Tuzla, Tuzla, School of Medicine, Bosnia and Herzegovina)
  • Mirela Delibegović (University Clinical Centre Tuzla, Tuzla, Clinic for Radiology, Bosnia and Herzegovina)
  • Muhamed Katica (University of Sarajevo, Sarajevo, Veterinary Faculty, Department of Clinical Sciences of Veterinary Medicine, Bosnia and Herzegovina)
  • Muamer Obhodžaš (University of Sarajevo, Sarajevo, Veterinary Faculty, Department of Clinical Sciences of Veterinary Medicine, Bosnia and Herzegovina)
  • Muhamed Ođuz (Veterinary Station “Mak-vet d.o.o. Breza”, , Breza, Bosnia and Herzegovina)

Abstract

Aim: Clips in neurosurgery are made of titanium alloys, which reduce artifacts on computed tomography (CT). The radiological advantage of plastic clips on the CT image was demonstrated when they were placed in an inter-hemispherical position at an angle of 90º. The aim of this study was to investigate the behaviour of the clip placed at different angles.

Methods: Sixty heads of domestic pigs were divided into two groups, in group 1 a titanium clip was placed to the interhemispheric position at an angle of 90º, 45º, 0º, ten heads for each angle. In group 2 a plastic clip was placed in the same way. CT scan of the brain was performed for each angle. The size of the density and possible artifact were measured on CT.

Results: The size of the titanium clip ranged from 17.05 mm at an angle of 0º in the axial plane to 91.47 mm at an angle of 0º in the sagittal plane. The average size of the plastic clip ranged from 6.4 mm at an angle of 0º in the axial plane to 23.22 mm in an angle of 90º in the sagittal plane. Artifacts were observed only in the titanium clip.

Conclusion: Plastic clips have shown radiological advantages over titanium clips in the CT image. The average density size of the plastic clip in all planes and all angles was smaller than the titanium clip.

Keywords: aneurysm clipping, artifacts, computed tomography