Effects of post-sampling analysis time, type of blood samples and collection tubes on values of blood gas testing

By
Jasmina Smajić ,
Jasmina Smajić
Contact Jasmina Smajić

Clinical Chemistry and Biochemistry, Clinical Centre University of Sarajevo , Sarajevo , Bosnia and Herzegovina

Damira Kadić ,
Damira Kadić

Department of Laboratory Diagnostics, Cantonal Hospital Zenica , Zenica , Bosnia and Herzegovina

Sabaheta Hasić ,
Sabaheta Hasić

Department of Medical Biochemistry, School of Medicine, University of Sarajevo , Sarajevo , Bosnia and Herzegovina

Nafija Serdarević
Nafija Serdarević

Clinical Chemistry and Biochemistry, Clinical Centre University of Sarajevo , Sarajevo , Bosnia and Herzegovina

Abstract

Aim
To investigate effects of post-sampling analysis time, a type of blood samples and collection tubes on blood gas testing.
Methods
This study included 100 patients at the Clinic for Pulmonary Diseases, Clinical Centre Sarajevo. The partial pressure of
oxygen (pO2) and carbon dioxide (pCO2), and the oxygen saturation level of hemoglobin (sO2) were analyzed in the arterial bloodsamples (ABS) and capillary blood samples (CBS) by a potentiometric method using a blood gas analyzer ABL 555 (Radiometer, Copenhagen, Denmark). Paired measurements of ABS were performed within 15 minutes and after 60 minutes from sampling and compared. The results of CBS obtained within 15 minutes were compared with matching ABS results, as well as the results obtained from CBS within 15 minutes taken into glass and plastic tubes.
Results
pO2 and sO2 values were significantly lower after 60 minutes compared to those within 15 minutes in ABS (9.20±1.89 vs.
9.51±1.95 and 91.25±5.03 vs. 92.40±4.5; p<0.01, respectively). Values of pO2 and sO2 in CBS were significantly lower than values obtained in ABS (8.92±2.07 vs. 9.51±1.95 and 91.25±4.86 vs. 92.40±4.50; p<0.01, respectively). Obtained pO2 and sO2 values in CBS in the plastic tubes were higher than those in the glass tubes (8.50±1.98 vs. 7.89±2.0 and 89.66±11.04 vs. 88.23±11.22, p<0.01 respectively). pCO2 blood values were not influenced significantly (p>0.05).
Conclusion
The length of post-sampling analysis time, a type of blood samples and collection tubes have significant impact on
blood oxygen parameters. Analysis within 15 minutes after blood sampling is considered as appropriate. 

References

1.
Baird G. Preanalitical considerations in blood gas analysis. Biochem Med (Zagreb). 2013;19–27.
2.
Clsi. Blood Gas and pH Analysis and Related Measurements; Approved Guideline -second edition. CLSI Document C46-A2. 2010;
3.
Burnett R, Covington A, Fogh-Andersen N, Külpmann W, Maas A, Müller-Plathe O, et al. Scientific Division. Committee on pH, Blood Gases and Electrolytes. Approved IFCC recommendations on whole blood sampling, transport and storage for simultaneous determination of pH, blood gases and electrolytes. Eur J Clin Chem Clin Biochem. 1995;247–53.
4.
Clsi. Procedures and Devices for the Collection of Diagnostic Capillary Blood Specimens; Approved Standard -sixth edition. CLSI Document H04-A6. 2008;
5.
Hardinge M, Annandale J, Bourne S, Cooper B, Evans A, Freeman D, et al. British Thoracic Society Home Oxygen Guideline Development Group; British Thoracic Society Standards of Care Committee. British Thoracic Society guidelines for home oxygen use in adults. Thorax. 2015;1–43.
6.
Srisan P, Udomsri T, Jetanachai P, Lochindarat S, Kanjanapattanakul W. Effects of temperature and time delay on arterial blood gas and electrolyte measurements. J Med Assoc Thai. 2011;9–14.
7.
Beaulieu M, Lapointe Y, Vinet B. Stability of PO 2 , PCO 2 , and pH in fresh blood samples stored in plastic syringe with low heparin in relation to various blood-gas and hematological parameters. Clin Biochem. 1999;101–7.
8.
Knowles T, Mullin R, Hunter J, Douce F. Effects of syringe material, sample storage time, and temperature on blood gases and oxygen saturation in arterialized human blood samples. Respir Care. 2006;732–6.
9.
Mohammadhoseini E, Safavi E, Seifi S, Seifirad S, Firoozbakhsh S, Peiman S. Effect of Sample Storage Temperature and Time Delay on Blood Gases, Bicarbonate and pH in Human Arterial Blood Samples. Iran Red Crescent Med J. 2015;13577.
10.
Zavorsky G, Lands L, Schneider W, Carli F. Comparison of fingertip to arterial blood samples at rest and during exercise. Clin J Sport Med. 2005;263–70.
11.
Murphy R, Thethy S, Raby S, Beckley J, Terrace J, Fiddler C, et al. Capillary blood gases in acute exacerbations of COPD. Respir Med. 2006;682–6.
12.
Zavorsky G, Cao J, Mayo N, Gabbay R, Murias J. Arterial versus capillary blood gases: a meta analysis. Respir Physiol Neurobiol. 2007;268–79.
13.
Smeenk F, Janssen J, Arends B, Harff G, Van Den Bosch J, Schönberger J, et al. Effects of four different methods of sampling arterial blood and storage time on gas tensions and shunt calculation in the 100% oxygen test. Eur Respir J. 1997;910–3.
14.
Schmidt C, Müller-Plathe O. Stability of pO2, pCO2 and pH in heparinized whole blood samples: influence of storage temperature with regard to leukocyte count and syringe material. Eur J Clin Chem Clin Biochem. 1992;767–73.
15.
Smajić1 J, Kadić2 D, Hasić3 S. Nafija Serdarević1 1Klinička hemija i biohemija, Klinički centar Univerziteta u Sarajevu, Sarajevo, 2Služba za laboratorijsku dijagnostiku.
16.
Zenica Z. Medicinski fakultet Univerziteta u Sarajevu.

Citation

Authors retain copyright. This work is licensed under a Creative Commons Attribution 4.0 International License. Creative Commons License

 

Article metrics

Google scholar: See link

The statements, opinions and data contained in the journal are solely those of the individual authors and contributors and not of the publisher and the editor(s). We stay neutral with regard to jurisdictional claims in published maps and institutional affiliations.